Notes on Attached Software

l

Before opening the media package, read YFGW710 Software License Agree-

ment. Once the package is opened, it is thought that the contract is agreed.

l

Yokogawa assumes no responsibilities for this software except as stated in

the warranty.

l

This software is used with this product exclusively. This can be installed in the

computer which connects with this product.

n

Insulation Resistance and Dielectric Strength Test

l

Insulation Resistance and dielectric strength test could not be carried out

for YFGW710, due to the circuit ground is connected directly to the frame of

YFGW710 via RF cable shield.

n

Cautions for Safely Applying the Device

l

EMC Conformity Standards

EN61326-1 Class A, Table 2 (For use in industrial locations), EN61000-6-2

CAUTION

This instrument is a Class A product, and it is designed for use in the industrial environment.
Please use this instrument in the industrial environment only.

l

R&TTE Conformity Standards:

ETSI EN 300 328, ETSI EN 301 489-17, EN61010-1, EN62311
• Indoor/Outdoor use

l

Low Voltage Directive

Applicable standard: EN61010-1
(1) Pollution Degree 2
“Pollution degree” describes the degree to which a solid, liquid, or gas which deteriorates dielec-

tric strength or surface resistivity is adhering. “2” applies to normal indoor atmosphere. Normally,

only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by

condensation must be expected.
(2) Installation Category I
“Overvoltage category (Installation category)” describes a number which defines a transient over-

voltage condition. It implies the regulation for impulse withstand voltage.
“I” applies to electrical equipment which is supplied from the circuit when appropriate transient

overvoltage control means (interface) are provided.