Transmitting 12 Lead ECG Patient Records
You can transmit 12 Lead ECG patient records to a remote handheld device running ZOLL
Data Relay software without placing the M Series unit into Realtime mode.
Procedure
Follow these steps to transmit patient record data to a remote handheld device.
2.
Press the Data Relay On softkey.
The M Series unit activates automatic transmission and
displays the Realtime mode indicator in the upper left
hand corner of the screen.
The Bluetooth module’s green power LED illuminates.
The M Series Bluetooth module can now establish a
connection to the remote handheld device, and will
automatically transmit 12 Lead ECG and trend data.
Step Action
Result
CHARGER ON
SUMMARY
CODE
MARKER
NIBP
00:17
MONITOR
ECG
24
NIBP
100
120
80
SpO2%
98
CO2
38
m
m
Hg
24
RR
ECG
II
120
x2
LEAD
SIZE
HR
**
Param
Wave 2
Alarms
12 Lead
ID#
Realtime
mode
indicator
Step Action
Result
1.
With the M Series unit in Monitor
mode (ON for AED units), press the
12 Lead softkey.
The M Series unit displays the 12 Lead menu.
2.
Press the PT Info softkey.
The M Series unit displays the Patient Information menu.
CHARGER ON
SUMMARY
CODE
MARKER
NIBP
01:20
12 LEAD MONITOR
ECG
24
NIBP
100
120
80
SpO2%
98
CO2
38
m
m
Hg
24
RR
ECG
II
120
x2
LEAD
SIZE
HR
Acquire
Settings
Alarms
Return
PT Info
CHARGER ON
SUMMARY
CODE
MARKER
NIBP
01:22
24
NIBP
100
120
80
SpO2%
98
CO2
38
m
m
Hg
24
RR
ECG
II
120
x2
LEAD
SIZE
HR
ID #
Age
Patient
Records
Return
Gender
PATIENT INFORMATION
:
Press Return to Exit to
12 Lead Monitor Mode
60
Male