ZOLL M Series Defibrillator Rev G NIBP User Manual
Page 30
N
ON
-I
NVASIVE
B
LOOD
P
RESSURE
(NIBP)
24
ZOLL M Series Operator’s Guide - Option Insert
* Neonatal and pediatric NIBP modes available on M Series CCT units only.
Table 11 Battery Operating Time Specifications
For a new, fully charged battery
pack at 20°C with NIBP, EtCO
2
and
SpO
2
Options
35 defibrillator discharges at maximum energy (360J), or 1.25 hours
minimum of continuous ECG monitoring and blood pressure
measurements once every 5 minutes, or 0.75 hour of continuous ECG
monitoring/pacing at 60 mA, 70 beats/min.
Table 12 Environmental Specifications
Operating Temperature
0
° to 50° C
Storage and Shipping Temperature
-20
° to 70° C
Electromagnetic Immunity
AAMI DF-2: 1996, EN 61000-4-3: 1996, 20 V/m
Humidity
5 to 95% relative humidity, non-condensing
Operating Pressure
594 to 1060 mBar
Table 13 Alarm Limits Specifications
Systolic
Adult
High 80 – 260 mmHg;
Low 40 – 140 mmHg;
High 10.7 – 34.7 kPa
Low 5.3 – 18.7 kPa
Pediatric*
High 80 – 160 mmHg;
Low 35 – 140 mmHg;
High 10.7 – 21.3 kPa
Low 4.7 – 18.7 kPa
Neonatal*
High 60 – 130 mmHg;
Low 25 – 120 mmHg;
High 8.0 – 17.3 kPa
Low 3.3 – 16.0 kPa
Diastolic
Adult
High 50 – 200 mmHg;
Low 25 – 100 mmHg;
High 6.7 – 26.7 kPa
Low 3.3 – 13.3 kPa
Pediatric*
High 50 – 130 mmHg;
Low 20 – 100 mmHg;
High 6.7 – 17.3 kPa
Low 2.7 – 13.3 kPa
Neonatal*
High 30 – 105 mmHg;
Low 20 – 100 mmHg;
High 4.0 – 14.0 kPa
Low 2.7 – 13.3 kPa
Mean
Adult
High 60 – 220 mmHg;
Low 30 – 120 mmHg;
High 8.0 – 29.3 kPa
Low 4.0 – 16.0 kPa
Pediatric*
High 60 – 140 mmHg;
Low 20 – 120 mmHg;
High 8.0 – 18.7 kPa
Low 2.7 – 16.0 kPa
Neonatal*
High 35 –110 mmHg;
Low 20 – 105 mmHg;
High 4.7 – 14.7 kPa
Low 2.7 – 14.0 kPa