beautypg.com

Advisory and semi-automatic modes – ZOLL M Series Defibrillator Rev YC User Manual

Page 31

background image

Configuration Option Tables

27

Advisory and Semi-Automatic Modes

Feature

Options

Default

Energy Level: Shock 1

Damped Sine Wave (DSW): 1-10, 15, 20, 30, 50,
75, 100, 150, 200, 300, 360 J

Biphasic: 1-10, 15, 20, 30, 50, 75, 100, 120, 150,
200 J

Manual DSW: 200 J
Manual Biphasic: 120 J

AED DSW: 200 J
AED Biphasic: 120 J

Energy Level: Shock 2

Damped Sine Wave (DSW): 1-10, 15, 20, 30, 50,
75, 100, 150, 200, 300, 360 J

Biphasic: 1-10, 15, 20, 30, 50, 75, 100, 120, 150,
200 J

Manual DSW: 300 J
Manual Biphasic: 150 J

AED DSW: 200 J
AED Biphasic: 120 J

Energy Level: Shock 3

Damped Sine Wave (DSW): 1-10, 15, 20, 30, 50,
75, 100, 150, 200, 300, 360 J

Biphasic: 1-10, 15, 20, 30, 50, 75, 100, 120, 150,
200 J

Manual DSW: 360 J
Manual Biphasic: 200 J

AED DSW: 360 J
AED Biphasic: 200 J

Auto Analyze

Yes, No, Only in Semi-Auto

Manual: No

AED: Only in Semi-Auto

Stacked Shocks

1, 3

3

Display “DO CPR” Message

Yes, No, Only in Semi-Auto

Manual: No

AED: Only in Semi-Auto

CPR Message After Shock

“IF NO PULSE, PERFORM CPR;” “PERFORM
CPR”

“IF NO PULSE,
PERFORM CPR”

CPR Message After No Shock Advised

“IF NO PULSE, PERFORM CPR;” “PERFORM
CPR”

“IF NO PULSE,
PERFORM CPR”

Duration of CPR Interval

1, 1.5, 2, 2.5, 3, 3.5 or 4 minutes

1 minute

Record CPR Compression Data

Yes, No

No

CPR Compression Metronome

Yes, Semi-Auto Only, No

Semi-Auto Only

CPR Data Displayed

R/D Indicators, Rate Only

R/D Indicators

Check Pulse

Yes, No, After Shock Only, After No Shock Only

Yes

Restart Analysis After CPR

Yes, No

Yes

“CHECK PATIENT” Prompt

“CHECK PATIENT,” “PRESS ANALYZE”

“CHECK PATIENT”

Display ECG in Semi-Auto Mode

Yes/No

Yes

Store to Card in Auto Monitor Mode

Yes/No

Yes

Auto Charge in Advisory Mode

Yes/No

Yes

ECG Waveform to Card in Advisory Mode

Yes/No

No

Audio Data Recording to Card in Advisory
mode

Yes/No

No

Enable Voice in Advisory Mode

Yes/No

No

Display Heart Rate in Semi-Auto Mode

Yes/No

No

Auto-Analyze on Power Up

Yes/No

No

Extra CPR Interval Before 1st Analysis

Yes/No

No

Duration of Extra CPR Interval

1, 1.5, 2, 2.5, 3, 3.5, or 4 minutes

1 minute

Enable 12-Lead in Semi-Auto Mode

Yes/No

Yes

AED Defib Ready Hold Time

15 seconds, 60 seconds

15 seconds