Press the Change Value softkey to scroll through the available values for each parameter.

Press the Next Item softkey to move to the next parameter in the table.

Press the Return softkey to save the values entered and return to the main configuration menu.

You can set the values for the following Clock Synchronization options:

Time Zone:

Selects the time zone in which the M Series will be used. Choices
include: Azores, Middle Atlantic, S. America Eastern, AST Atlantic,
EST Eastern, CST Central, MST Mountain, PST Pacific, AKST
Alaska, HST Hawaiian, Samoa, Dateline, GMT Greenwich, CET
Central Europe, EET Eastern Europe, MSK Moscow, Arabian,
Afghanistan, West Asia, Central Asia, Southeast Asia, AWST
Australian West, Korean, ACST Australian Cent, AEST Australian
East, Central Pacific, or New Zealand

Dial Location:

Sets the National Institute of Science and Technology (NIST) dialing
location to either Hawaii or Colorado. This is the location called by
the M Series to obtain accurate time information for setting its internal
clock.

Prefix:

Affixes up to eight user selectable digits or special characters (+, &, !,
*, or comma) to the start of the dialing location phone number.

Daylight Savings

When set to “Yes” and the user dials the selected NIST location, the unit will automatically
adjust the time for Daylight Savings based on the assumption that in North America Daylight
Savings Time starts on the second Sunday of March at 1:59:59 AM (setting the time ahead by
one hour) and ends on the first Sunday of November at 1:59:59 AM (setting the time back by
one hour).

To enable this option, you must select a time zone within North America (for example, EST,
CST, MST, PST, or AKST). Otherwise, the unit disables the Daylight Savings option.

Card Capacity Message Enabled

When set to “Yes” and data recording is enabled with a data card present, the M Series will
display “CARD LOW” messages when the remaining storage capacity of the card reaches 30,
15, and 5 minutes.

To enable this option, the “Report Memory Card Errors” option must be set to “Yes”;
otherwise, the option is ignored.

Erase Card Prompt at Pwr Off

Sets the M Series unit to allow the user to erase the data card at power-off if the card contains
only self test, and optionally, synchronized cardioversion data. To enable this option, the
“Allow Card Erase” option must be set to “Yes”.

If set to “Yes”, the unit will display the message “Erase Card -- Yes/No?” when the user turns
the Selector Switch to the OFF position and the following criteria are met:

1. A data card is installed in the M Series unit.
2. The unit has NOT performed a defibrillation, other than a self test, since power-up.

Note:The M Series unit can also be configured to display the “Erase Card -- Yes/No?”

prompt if the card contains only synchronized cardioversion data. See the “Sync
Defib Excluded” option for more information.

3. The unit has NOT delivered a pacing current greater than 25 mA since power-up.

If the user selects “Yes” at the prompt within 15 seconds, the card will be erased and the unit
will subsequently shut off. If the user selects “No”, or 15 seconds is exceeded without a
response, the unit will shut off without erasing the card.