Specifications, General – ZOLL M Series Defibrillator Rev M User Manual
Page 14

M S
ERIES
O
PERATOR
’
S
G
UIDE
- O
PTION
I
NSERT
SpO
2
- 12
Specifications
Note: The M Series Pulse Oximetry Option is calibrated for Functional saturation.
General
Saturation (% SpO
2
) Range
1% - 100%
Pulse Rate (bpm) Range
25 - 240 bpm
Saturation (% SpO
2
) Accuracy
During No Motion Conditions
Adults
70% - 100%, ±2 digits
0% - 69%, unspecified
Neonates
70% - 100%, ±2 digits
0% - 69%, unspecified
Saturation (% SpO
2
) Accuracy
During Motion Conditions
Adults
70% - 100%, ±3 digits
0% - 69%, unspecified
Pulse (bpm) Accuracy
During No Motion Conditions
25 - 240, ±3 digits
Pulse (bpm) Accuracy
During Motion Conditions
25 - 240, ±5 digits
SpO
2
Alarm Limits
On/Off displayed on monitor. User selectable.
High 72 - 100% saturation, Low 70 - 98% saturation
Pulse Rate Alarm Limits
On/Off displayed on monitor. User selectable.
High 60 - 235 bpm, Low 20 - 100 bpm
Saturation (% SpO
2
) Resolution
1%
Pulse Rate Resolution
1 bpm
Text Display Update Rate
100 milliseconds
Trace Update Rate
52 milliseconds
Bio-Compatibility
Patient contacting material meets requirements of ISO 10993-1,
Biological Evaluation of Medical Device - Part I, for external
devices, intact surfaces and short-term exposure.
Environmental
Operating Temperature:
0° to 40° C
Storage and Shipping Temperature: -20° to 60° C
Note: The M Series device may not perform to specifications
when stored at the upper or lower extreme limits of storage
temperature and immediately put into use
Electromagnetic Immunity
(SpO
2
Option Only)
AAMI DF-2: IEC 1000-4-3 to 18 v/m
Operating Time
For a new, fully charged battery pack at 20°C:
•
35 defibrillator discharges at maximum energy (360J), or
•
2.0 hours minimum of continuous ECG monitoring, or
•
1.75 hours of continuous ECG monitoring/pacing at 60 mA,
70 beats/min.