Default settings, Accessories – ZOLL M Series Defibrillator Rev M User Manual
Page 12
![background image](/manuals/355854/12/background.png)
M S
ERIES
O
PERATOR
’
S
G
UIDE
- O
PTION
I
NSERT
SpO
2
- 10
Default Settings
When the pulse oximeter is turned on, the following default settings are automatically selected and remain in operation
until changed.
Note: Only the default settings are available in AED Semi-Automatic operation. Default alarm limit setting may be
adjusted in System Configuration mode. See the M Series Configuration Guide for more information.
SpO
2
Accessories
The following table describes each of the SpO
2
accessories.
Note: LNOP single use and reusable sensors are no longer available, but remain compatible with the M Series.
Parameter
Default Setting
Range
Averaging Mode
8 seconds (medium)
4 seconds (short)
8 seconds (medium)
16 seconds (long)
Sensitivity
Normal
Normal or High
High SpO
2
Alarm Limit
OFF (appears as: ---)
50% to 100% or OFF
Low SpO
2
Alarm Limit
85%
50% to 100% or OFF
High Heart Rate Alarm Limit
150 bpm (beats/minute)
60 to 280 bpm (monitoring via ECG)
60 to 235 bpm (monitoring via Pulse Oximetry)
Low Heart Rate Alarm Limit
30 bpm
20 to 100 bpm
Item
Description
Part #
LNCS Adtx
Single use sensor for patients > 30 kg
8000-0320
LNCS Pdtx
Single use sensor for Pediatrics and Slender Adults 10 - 50 kg
8000-0321
LNCS Inf-L
Single use sensor for Infants 3 - 20 kg
8000-0322
LNCS Neo-L
Single use sensor for Neonates < 3 kg
8000-0323
LNCS NeoPt-L
Single use sensor for Neonates < 1 kg (Pre-term)
8000-0324
LNCS DCI
Reusable sensor for Adults and Pediatrics > 30 kg
8000-0294
LNCS DCIP
Reusable sensor for Pediatrics 10 - 50 kg
8000-0295
LNC-4
4' Reusable Patient Cable
8000-0298
LNC-10
10’ Reusable Patient Cable
8000-0293
LNC Ext
LNC Extension Cable, DB-9 Termination, 4ft
8000-0325
LNCS-to-LNOP
Adapter Cable, LNCS Sensor to LNOP Patient Cable
8000-0327
LNCS DC-12
LNCS Adult Reusable Direct Connect 12’ Cable
8000-0296