Messages and troubleshooting – ZOLL M Series Defibrillator Rev M User Manual
Page 13
Pulse Oximetry (SpO
2
)
SpO
2
- 11
Messages and Troubleshooting
The following chart lists the messages that may appear on the M Series unit relating to SpO
2
, why the message
appeared, and the action(s) to take if the message indicates a problem.
The operator should become thoroughly familiar with this information before using the oximeter for patient monitoring.
Message
Possible Cause(s)
Recommended Action(s)
SpO2 AMBIENT LIGHT/
CHECK SpO2 SENSOR
Excessive ambient light.
Relocate the sensor to a site more
shielded from light or reduce the
amount of light shining on the sensor.
SpO2 PULSE SEARCH
Appears when the oximeter cannot
detect a patient's pulse.
Reposition or relocate the sensor, and/
or increase perfusion.
CHECK SpO2 SITE
Insufficient perfusion at sensor site.
Reposition or relocate the sensor, and/
or increase perfusion.
CHECK SpO2 SENSOR
Appears when SpO
2
readings may
be invalid due to motion, to an
unacceptable sensor site, to poor
placement, low perfusion, or sensor
is OFF.
For all causes, reposition or relocate the
sensor, and/or increase perfusion.
SpO2 FAULT XX
Appears when SpO
2
subsystem of
the unit has failed.
Call ZOLL Technical Service
Department.
Dashes (-----) appear in place of
SpO2 numeric and do not
change to a real number.
Excessive ambient light, inadequate
perfusion, high signal artifact, a
defective or disconnected sensor,
etc.
Reposition or relocate the sensor, and/
or increase perfusion.