ZOLL IVTM™ System
Physicians' Manual
600248-001 Rev 3
16
The reduction in fever burden was accompanied by a reduction in the use of
adjunctive cooling means as presented in the tables below.
Table 4. Use of Topical Cooling Devices
Cool Line
Control
% Reduction
p*
One or more topical cooling
device (n/N, %)
26 / 154
16.9
67 / 142
47.2
64%
<0.0001
Cooling Blanket use (n/N, %)
25 / 154
16.2
59 / 142
41.6
61%
<0.0001
Other device use (n/N, %)
7 / 154
4.6
19 / 142
13.4
66%
0.008
* Fisher’s exact test
36
36.5
37
37.5
38
38.5
39
0
20
40
60
Hours
Cool Line Patients (Mean Temperature ± SD)
36
36.5
37
37.5
38
38.5
39
0
20
40
60
Hours
Control Patients (Mean Temperature ± SD)