Vanguard PCI-600 User Manual
Page 6

PCI-600 USER’S MANUAL
REV 2
3
1.2
PCI-600 Technical Specifications
Table 1. PCI-600 Technical Specifications
TYPE 100 - 600 Amp current source
PHYSICAL SPECIFICATIONS Dimensions: 17”W x 12.5”H x 10.5”D (42.6 cm x 32.0 cm x 27.0 cm);
Weight: 46 lbs (21 kg)
INPUT POWER
100 – 120 Vac or 200 – 240 Vac (factory pre-set), 50/60 Hz
INTERNAL METER RANGE
100 mA – 1000 A; Accuracy: 1% of reading ±20 mA
MEASUREMENT METHOD
Isolated CT
EXTERNAL METER RANGE
10 mA – 10 A; Accuracy: 1% of reading, ±2mA
MEASUREMENT METHOD
Isolated CT
TIMER READING RANGE
1ms – 2 hours; Accuracy: 0.1% of reading ±1ms
TIMER STOP INPUTS
Voltage input (24V – 300V, DC or peak AC), dry contact input, or removal
of primary current
DISPLAY
Back-lit LCD Screen (20 characters by 4 lines); viewable in bright sunlight
and low-light levels
COMPUTER INTERFACE
RS-232C port for factory calibration and diagnostics
SAFETY
Designed to meet IEC61010 (1995), UL61010A-1, CSA-C22.2 standards
ENVIRONMENT
Operating: -10˚C to 50˚ C (15˚F to +122˚ F); Storage: -30˚ C to 70˚ C (-
22˚F to +158˚ F)
CABLES
10-foot #1/0 AWG test leads, power cord, ground cable
OPTIONS
Transportation case
WARRANTY
One year on parts and labor
NOTE
The above specifications are valid at nominal operating voltage and at a
temperature of 25
°C (77°F). Specifications may change without prior notice.