Qphy-most50e variables, Default bitrate, Fixed bitrate mode – Teledyne LeCroy QPHY-MOST50 User Manual

Page 17: Debug mode, Demo mode, Dut (device under test) timing mode, Clock recovery edge, Most root function, Absolute eye masks size, Most minus signal input

background image

QPHY-MOST50e Software Option

920457 Rev B

17

QPHY-MOST50E VARIABLES

Default Bitrate

Specifies whether the bitrate is based on 48 kHz or 44.1 kHz frames.

Fixed Bitrate Mode

Specifies whether the fixed bitrate defined above is used instead of the measured value.

Debug Mode

When active, the test will pause after each test for debugging purposes and displays the parameter
values.

Demo Mode

Specifies whether to use saved waveform data for testing or captured live signals.

DUT (Device Under Test) Timing Mode

Specifies the type of the DUT: Master or Slave.

Clock Recovery Edge

Specifies the clock edge use for Eye diagrams. Positive is the default and is required for compliance.

MOST Root Function

Specifies the math function to be used for the tests. Where additional function channels are required they
are allocated consecutively from the set value. A minimum of 3 function channels are required.

Absolute Eye Masks Size

Specifies whether the scaling of absolute eye diagrams is determined by the eye mask or the waveform
amplitude.

MOST Minus Signal Input

Specifies the input channel for the negative signal when using a single ended probe or direct connections
(SMA cables). Set to 'None' for compliance testing using a differential probe.

MOST Plus Signal Input

Specifies the differential probe channel or the input channel for the positive signal when using direct
connections (SMA cables).

MOST Parameter

Specifies the first available parameter to be used for the tests. Where additional parameters are required
they are allocated consecutively from the set value. A minimum of 3 parameters are required.

Quick Mode

Specifies if existing channel settings are reused for tests in the same group where possible. This is the
default.

Input Row For Relevant Zi Models

Specifies whether to use the A or B inputs when a Zi model oscilloscope with multiple input rows is
detected. Row A (Top Input Row) is the default.