JUMPER TABLE

Description

Jumper

Utilization

Current measurement

J3

Outputs to DVM

External power input

J5

Inputs from external power supply

+5V isolation from the system

JP5

On = voltage from system
Off = voltage from External supply

+12V isolation from the system

JP6

On = voltage from system
Off = voltage from External supply

VCC3 isolation from the system

JP7

On = voltage from system
Off = voltage from External supply

VDDQ isolation from the system

JP9

On = voltage from system
Off = voltage from External supply

Disable Mechanical On-Off Switch

SW

On = S1 can control the board
Off = S1 is disabled

Remote On-Off

JP15

Shunt On, Board Id is One
Shunt Off, Board Id is Two

JP16

Factory Use Only

LEDs

There are two sets of LED on the board, green and red. Each LED has a label which Voltage it represents. The green
LED, when lit, indicates presence of the corresponding voltage. The red LED, when lit, represent a short on the
corresponding voltage. Please note that the green LEDs do not indicate if the voltage is at the expected level or not.