Operator’s Manual
922209-00 Rev A
3
Do not overload. To avoid electric shock or fire, do not apply any potential to the
probe leads that exceeds the maximum rating of the probe.
Comply with voltage derating curve. When measuring higher frequency signals,
comply with the Voltage vs. Frequency Derating Curve.
Observe all terminal ratings. To avoid electric shock or fire, observe all markings on
the oscilloscope before connecting. Consult the respective oscilloscope product
manual for further ratings information.
Do not remove probe casing. Removing the probe’s case or touching exposed
connections may result in electric shock.
Use only within operational environment listed. Do not use in wet or explosive
atmospheres. Keep product surfaces clean and dry.
Use only accessories compatible with the probe.
Handle with care. The probe tip is extremely sharp and may puncture skin or cause
other bodily injury if not handled properly.
Keep fingers behind the finger guard of probe body and accessories.
Do not operate with suspected failures. Before each use, inspect the probe and
accessories for any potential damage such as tears or other defects in the probe
body, cable jacket, accessories, etc. If any part is damaged, cease operation
immediately and sequester the probe from inadvertent use.
Operating Environment
The accessory is intended for indoor use and should be operated in a clean, dry
environment. Before using this product, ensure that its operating environment is
maintained within these parameters:
Temperature: Operating, 5° to 40° C; Non-operating,
-
20° to 60° C
Humidity:
5% to 85% relative humidity (%RH) up to +30° C
5% to 65% RH above +30° C to 40° C
5% to 45% RH above 40° C
Altitude:
Up to 3000 m (9842 ft)