PP007 Passive Probe

4

922207-00 Rev A

Introduction

The PP007 is a miniature high impedance passive probe. Its high input resistance
and low capacitance make it ideal from general purpose probing of signals with
frequency content from DC through several hundred MHz. The PP007 has a
large selection of connection accessories, both supplied standard with the probe
and available from Teledyne Lecroy as optional accessories.

The PP007 is available in two models, differing only in the design of the
compensation network.

The PP007-WR is designed for use with Teledyne Lecroy WaveRunner series
oscilloscopes.

The PP007-WS is designed for use with Teledyne Lecroy WaveSurfer series
oscilloscopes.

All standard and optional accessories are interchangeable between all models.

Specifications

Electrical Characteristics

Attenuation

10

1%

Input Resistance

10 M

1%

Input Capacitance

9.5 pF

Input Impedance

(see plot on page 5.)

Compensation Range

10 – 20 pF

Bandwidth

500 MHz (-3 dB)

Electrical Ratings*

Maximum Input Voltage

Measurement category I: 400 V rms,

1250 V transient overvoltage

(see voltage derating curve on next page)

Measurement category II: 300 V rms CAT II

Pollution Degree

2