Low Flow Applications (5 to 500 ml/min)
Using Multiple-tube Adjustable Low Flow Holder
Setup
For a diagram of the pump, see Figure 1, page 5.
Figure 3
Quad Adjustable Low Flow Holder
1 Anti-tamper cover
2 Flow adjust screws
3 Rubber sleeve
4 Sorbent sample tube
5 Protective cover
Top view of fl ow
adjust screws
1
Charger and battery
pack connected
Install battery (see Installing the Battery Pack on page 21). For op-
timum charge, ensure pump is not running.
Charge the battery by
connecting the charger plug to the sampler charging jack (Figure
1, #22). Ensure that the battery is fully charged before sampling.
After charging the battery pack, it is good practice to run the pump for
approximately fi ve minutes before calibrating. This ensures the battery is in
more steady-state conditions and improves the agreement in pre and post-
sampling calibrations.
Do not charge or operate pump from charger in
hazardous locations.
Use only an SKC-approved charger designat-
ed for this model to ensure reliable perfor-
mance. Failure to do so voids any warranty.
Ensure proper orientation of charge cable
before plugging it into the charging jack.
Improper orientation/contact will short-
circuit the battery and voids any warranty.
Short-circuiting the battery pack will
render it immediately inoperative.
Failure to follow warnings and cautions
voids any warranty.
The battery pack may be kept on the SKC-
approved charger for an indefi nite time.