Maintenance

Sample Hose Inspection and Cleaning

Inspection Frequency: Inspect sample hoses before and after each use of the instrument

and after every replacement of the sample filter.

What to Look For: Inspect the inside diameter of the inner tube for signs of oil buildup,

dirt, obstructions, kinks, cuts, fraying, or any other signs of wear. Oil contamination can

be cleaned out of the sample hoses as directed below. Hoses that show signs of wear

should be replaced immediately to avoid danger of rupture or bursting.

Sample Hose Cleaning Procedure:

1. Remove the hose from the instrument and flush it with isopropyl alcohol or methanol

until the oil is thoroughly cleansed from the inner tube.

WARNING

To prevent personal injury, flush hoses away from sparks, open flames
or other ignition sources and in an area that is well ventilated.

2. Dry the hose by blowing clean, dry nitrogen or shop air through the inner hose, or by

allowing the hose to air-dry for several hours. Do not dry the hose with shop air that is

lubricated.

3. When the hose is completely dry, reinspect the hose for signs of wear, as described

above, and replace the hose if wear is evident.