Operating procedures – Robinair 16910 ID Plus Refrigerant Identifier User Manual
Page 14
12
Operating Procedures
C. When the correct setting of local elevation has been achieved, allow the instrument to
sit for 20 seconds while not pushing either of the buttons. The instrument automatically
returns to the warm-up period, and the elevation setting is stored in memory.
5. After completion of the warm-up period, the instrument self-calibrates. Ambient
air is drawn in through the air intake port and is presented to the sensing device for
calibration. The calibration period runs for 20 seconds.
6. When the calibration period is completed, the instrument displays READY: CON.
HOSE, PRESS A TO START, and the green LED flashes. Connect the service end of
the sample hose to the low side or vapor port of the refrigerant storage vessel or vehicle
air conditioning system to be sampled. When the hose is secured into position, press
the “A” button of the instrument to start processing. NOTE: For correct operation, the
instrument pressure gauge should read at least 10 psig.
Tech Tip: To achieve the best air detection accuracy, allow the instrument to warm up
an additional 5 minutes before pressing the “A” button as stated in Step 5. This allows
additional warming and stabilization of the sensing device before sampling. During
the additional warm-up period, the instrument will “Time-Out” and display RECAL.
Disconnect the sample hose from the refrigerant source, and press the “A” button to
re-calibrate. Once re-calibration is complete, it is not necessary to repeat additional
warm-up periods as long as the instrument is not powered down between sample runs.
7. The instrument reads SAMPLING IN PROGRESS while a small refrigerant sample
is analyzed to determine the concentrations of R-12, R-134a, R-22, hydrocarbons, and
air. Once analysis is complete, the percentage of concentrations of R-12, R-134a, R-22,
and hydrocarbons is displayed. Press “A” to print the results, press “B” to exit (if there
is no air), or press “B” to continue the purge function if air is present.