Error Codes
Error Codes and Corrective Actions
The instrument is supplied with self-diagnostic software to guide you through problems
that may occur in the field. The software supples three codes that direct you toward
corrective measures. Should the corrective measures not clear the problem, contact
Technical Services for assistance.
ERR.1
Signifies unstable readings due to inconsistent sample supply or interference from
strong electrical fields.
Corrective Action for Error 1:
1. If the error code occurs during a sampling mode, verify the sample hose is receiving a
sample of at least 9 psig. Also, verify the sample exhaust and air intake ports are clear
and not obstructed in any manner.
2. If this error occurs during a calibration mode, verify the purge vent hose is not directed
towards the instrument, and the air intake and sample exhaust ports are clear and not
obstructed.
3. The unit must be kept away from sources of strong electrical fields (such as large
compressors). Move the instrument 3 to 5 feet (1 to 1.5 meters) away from such
sources and try again.
4. If the instrument is being used in an environment outside the operating temperature
range of 39° F to 122° F (4° C to 50° C), allow the instrument an additional warm-up
period of 15 minutes after power-up before use.
5. To clear the error code, unplug the instrument. The error code will be cleared at the
next power-up.