beautypg.com

Ransburg Solo Installation & Set-Up Kit User Manual

Page 2

background image

SI-08-06.1

Below is a detailed description and proper usage

instructions for each item in the kit:

Items 1 and 2 - Cap Pressure Test Gauges and

Test Air Caps

The cap pressure test gauges (Item 1) are used

along with the test air caps (Item 2) to measure

the fan and atomization air pressures at the air

cap. To use the cap pressure test gauges and

test air caps, follow these steps:

1. Hook the applicator to main air. Turn main

air on.

2. Remove the retaining ring and air cap from

the applicator.

3. Place one of the test air caps (Item 2) onto the

applicator. To determine which test air cap to

use, look at the original air cap that was just

removed from the applicator. On one of the

faces, it will say 65V, 98V, or 122V. Select the

test air cap with the same numbers as on the

air cap.

4. Put the retaining ring back onto the applicator

and tighten.

5. Determine which gauge measures the

atomization air cap pressure. This will be the

gauge located above the word "atom" on the

gauge mounting bracket.

6. Next, determine which barbed fitting on the air

cap is the atomization air barb. This will be the

one located closest to the center of the air cap.

7. Place the tubing coming out from directly

underneath the atomization air pressure

gauge onto the atomization air barb on the air

cap.

Figure 2: Locate Atomization Air Pressure Gauge On Item

2

Figure 3: Locate Atomization Air Barb Fitting On Test Air

Cap

Figure 4: Connect Atomization Air Gauge to Atomization

Air Fitting

79651-00 Installation and Setup Kit