beautypg.com

Gentec-EO PC-MAESTRO (Old Software for MAESTRO) User Manual

Page 8

background image

PC-

MAESTRO User’s Manual Revision 1.0

7

SOFTWARE
TRIGGER LEVEL

0.1% to 99.9%, 0.1% resolution

Transfer Rate

No missing points throughput serial frequency : 300 Hz

For higher repetition rate, please use directly the MAESTRO

MAESTRO specification:

3

- 2 kHz for data acquisition in the real time full screen display, no missing

point

- 1.3 kHz for data acquisition in real time with time stamp, no missing

point

- 2 kHz in statistics mode, no missing point

From 2 kHz to 10kHz the MAESTRO will send the latest energy pulse

every 500µs. (corresponding to a subsampling at 2 kHz).

Frequency
Measurement

Accurate frequency measurement up to 14kHz

0-1000 Hz: 0.1% accuracy

1000-2000 Hz: 0.5%

2000-14000 Hz: 1%

Statistics

Current value, Max, Min, Average, Std Dev., RMS stability, PTP stability,

Time, Pulse #.

General Specifications

Serial
Communication

USB

RS-232

NOT AVAILABLE : Ethernet

Data Displays

Real Time, Scope, Averaging, Statistics, Digital Tuning Needle,

Histogram

User input correction
factors

1 multiplier and 1 offset (7 digits floating point)

Internet Upgrades

Downloadable link

No missing point
throughput serial
frequency (power
and energy)

300 Hz

Size

4.39 MB

Average RAM
allocation

9 to 23 MB

PC Recommended
Requirement

Dual Core Processor

3 Gb of RAM

Operating System
Compatibility

Windows 7

Windows Vista

Windows XP (not officially supported)

3