Set measure” options panel, Easure panel – Gentec-EO PC-MAESTRO (Old Software for MAESTRO) User Manual
Page 18

PC-
MAESTRO User’s Manual Revision 1.0
17
4.3.1
”Set Measure” Options Panel
Fig. 4-1 The Set Measure panel
Wavelength: Enter the desired wavelength. Press Enter or Tab to make the change effective.
Range: Use the dropdown menu to select the desired range. The chosen value will be the
maximum that the PC-MAESTRO will measure before displaying
“OUT”. By default, the
MAESTRO is in
“Auto” mode. For more information about the Auto Mode, please refer to the
MAESTRO user manual.
Measure Mode: Use the dropdown menu to set the measurement mode (Power, Single Shot
Energy, Moving Average, etc.). Please note that it can take up to 3 seconds for the PC-
MAESTRO to switch between measurement modes.
“Anticipation ON/OFF” and “Attenuator
ON/OFF
” can be controlled in this menu. Please refer to the MAESTRO user manual to get
more information about these functions.
Corrections: Use these settings to set a multiplier and/or an offset to your measurement. The
equation used is : Result = (Value*Multiplier)+Offset
Trigger Level:
Adjust the trigger level in 0.1% steps, from 0.1% to 99.9%.
Trigger level is not available in power mode
External trigger option is not available in power mode
Default value is 2%
For more information about the trigger level functionality, please refer to
the MAESTRO user manual.