The meter can be calibrated on the
humidity by end users with 33% and
75% salt solution. The ambient
condition is recommended to be at

o

23+/-2 C and stable humidity. Users
can terminate the calibration anytime
by turning off the meter.

RH CALIBRATION

6. TRIPOD MOUNTING

The meter can fit on a tripod for long

time monitoring. A built-in thread is
on the rear side of the meter.

5. AUTO POWER OFF

The meter turns off automatically
after 20 minutes non-operation.
To override the function, hold down
and for 2 seconds to turn on
the meter. The LCD will display “n”
and then enter measuring mode.
(Fig. C)