LOW ALARM: when the gas concentration is between the low
alarm and high alarm point, the detector will give a sound alarm
every 0.5 second and the yellow LED will light.
HIGH ALARM: when the gas concentration is above the high
alarm point, the detector will give a sound alarm every
0.5 second, the red LED and the yellow LED will light.
LOW VOLTAGE: when the battery power is lower than 2.7V, the
detector will give a sound alarm every 30 seconds.
SENSOR FAULT: when the sensor faults, the detector will give a
sound alarm every 1.5 seconds and the yellow LED will light.

6. NOTICES:

• The alarm may be activated if there is smoke, alcohol, perfume or

other volatile items in the vacinity.

• Please activate the product in clean air. If the detector is

activated in a gas concentration above 30%LEL, the detector may
indicate sensor fault. Then you need to reactivate the detector in
clean air.

• Avoid excess shock. Do not place any pressure on the tip.
• Low voltage may affect the sensitivity. Please replace the

batteries as soon as possible when the power is low.

• Don’t store the detector in the environment with caustic gas.
• Please operate according to the operation manual.