beautypg.com

Operating instructions – General Tools and Instruments DSM403SD User Manual

Page 7

background image

3. Press the SET button on the DSM403SD (callout 1-7 of Fig. 1), which has the term

“Fast/Slow” stenciled below it. If the word “FAST” is displayed, move ahead to Step 4. If
the word “SLOW” is displayed, press the SET button again to change “SLOW” to “FAST”.

4. Press the ▼ button on the DSM403SD (which has the term “A/C” stenciled above it). If

the letter “A” is displayed in the upper left corner, move ahead to Step 5. If the letter “C”
is displayed, press the button again to change it to an “A”.

5. When the display shows the terms “A”, “FAST” and “50-100”, the instrument is

ready to be calibrated. To calibrate the DSM403SD, set the output of the sound
calibrator to 94 dB and carefully turn the calibration screw on the right side of the meter
(callout 1-11 of Fig. 1) until the display shows the following: “94 ± 0.2 dB”.

OPERATING INSTRUCTIONS

To prepare to make sound level measurements, power on the DSM403SD and make sure
an SD card is firmly seated in the socket on the bottom of the meter. The meter’s default
settings are autoranging on (enabling measurements from 30 to 130 dB), “A” frequency
weighting and fast time weighting. Accordingly, when the meter is powered on the display
will show the following terms: “A”, “FAST” and “AUTO”. Following are procedures for
changing each of these parameters to suit the application.

If you already know that the volume of the machinery or environment you are measuring is
within a limited range, you may want to exit autoranging mode (the meter’s default setting)
and manually enter and use a specific measurement range.

1. To use a specific measurement range, press the button (which has the word

“Range” stenciled above it) once. The display will then show the first of the three range
options: “30 – 80” (representing a full-scale range of 30 to 80 dB). Press the button
again and the display will show the second option: “50 – 100”. Press the button again
and the display will show the third option: “80 – 130”.

The shape of the “A” frequency weighting curve simulates the response of the human ear,
and is therefore the better choice for measuring the sound level of an environment. By
comparison, the “C” weighting curve is flatter, and is therefore better for measuring the
sound level of a piece of machinery.

2. To switch from “A” frequency weighting (the meter’s default setting) to “C”

weighting, press the button (which has the term “A/C” stenciled above it). The letter
“A” at the upper left of the display will change to “C”. To return to “A” weighting, press
the button again.

The third sound measurement parameter you can change is the meter’s response time.
Fast time weighting, with a response time of 200 ms, simulates the response time of the
human ear and is better for measuring the volume of singular events. Slow weighting, with
a response time of 500 ms, is a better choice when you wish to measure the average
sound level that an ongoing process (such as the vibration of a machine) produces over
time.

7