OPERATING PROCEDURE:

ACCESSORIES

Upon receiving the shipment, inspect the container and equipment for any signs of damage.
Verify that you have received all the items listed: Meter, Standard Solution 1413 µS/cm,
Battery (installed), Lanyard, Instruction Manual and Carrying Case.

PREPARATION

1. Remove the protective cap from the meter. Rinse the electrode with clean water and wipe it

dry.

2. Press

button to turn the meter power on.

Note: Don’t touch or wipe the surface of inner black sensor of the conductivity cell.

CALIBRATION

Dip the conductivity cell into the standard solution 1413 µS/cm. Stir gently and wait until the
display is stabilized. Press and hold

button to enter calibration mode until the icon

CAL

appears in the display, and flashes 1413 µS/cm. When the display stops flashing and indicates
“SA”, and then “End” while calibration is ending, it will return to measurement mode.
Note: (1) Calibration with a 12.88 mS/cm standard solution would be better for measuring high

conductivity solution.
(2) The icon

will appear automatically during calibration mode.

(3) “SA” will not appear if the calibration fails.
(4) If the reading is not 0 µS/cm while the meter is in the air and not dipped into any

solution, then calibrate it in the air to make the reading become 0 µS/cm.

MEASUREMENT

1. After calibration, rinse the conductivity cell with clean water and wipe it dry. Dip the

conductivity cell into sample solution to be measured. Stir gently and wait until a stable
reading can be obtained.

2. Press

button to select the desired measurement mode Conductivity, TDS or Salinity.

Note: (1) “_ _ _ _” will appear in the display when it is over the measuring range.

(2) The unit will auto-range to µS/cm or mS/cm, ppm or ppt.

FUNCTIONS MODE

1. Press

button to enter Hold function mode.

The icon

HOLD will appear, and the reading value in the display can be locked.

Press button again to return to Measurement mode.

2. Press and hold

button to enter the measuring Maximum and Minimum function mode.

Hold button until a flashing icon

MAX and MIN appears in the display. The value of maximum

and minimum will appear in the display while pressing the button lightly. To exit this mode,
press and hold

button until icon

MAX and MIN disappears, and returns to Measurement

mode.