Section 4: Programming Configuration
4.1. Gaining Access to the program mode:
There is a pushbutton installed on the internal printed circuit board assembly.
Press the CAL button, S1 on the PCB, accessed through hole in back.
4.2. Keys in Programming/Calibration Mode
Each program step has more than one choice. The [UNITS] key will toggle the
displayed value, the [ZERO] key will accept the current displayed value, and move
to the next step.
4.3. Programming Steps
4.3.1. First program step: Zero range -- The display will show USA or CAn.
• The zero range is set at 2% or 1.4 lbs when programmed for Canadian use
• The zero range is set at 100% or 70 lbs when programmed for USA use
Press the [UNITS] key to change, [ZERO] key to accept.
4.3.2. Second program step: Active Units -- Diamonds on the bottom of the display
will point to the currently enabled units. Selections available are lb only, kg only, or
lb/kg selectable.
Press the [UNITS] key to change, [ZERO] key to accept.
REMOTE
DISPLAY
PORT
AC ADAPTER
USB
OUTPUT
CABLE
CALIBRATION
PROGRAMMING
&
PUSHBUTTON
REAR VIEW