beautypg.com

Teledyne PEM-9002 - Digital flue gas analyzer User Manual

Page 43

background image

7. System Maintenance

Waste Gas Cleaning System

:

Attention:

Empty the condensate reservoir completely after each measuring

operation. Water residues within the measuring instrument will destroy the
pumps and sensors!
Damage of the filter and / or improperly fitted filter will greatly decrease o

eliminate the filter function and will eventually destroy pumps and sensors.

Plug-type elements
and flanges:

Storage:

See drawing on page 42.

Check the microfilter for contaminations and replace as necessary.

If the pump capacity is reduced, exchange the diaphragm filter.

Make sure that threaded parts are straight when placed on and tighten them
moderately. Ensure sufficient sealing by means of O-rings.

Remove any gas residues. Grease with Vaseline.

Store in a cool and dry environment at a temperature of approx. 60 °F (20 °C).

38