Page 21

background image

6.5 Draft Measurement

mbar

hPa

from the measuring program:

S1

Hold measured values

S3

Terminate draft measurement

Hold

Return

Hold

Return

Draft Measurem.

T.Gas

415.4 °F

Draft

0.00 InW

Before pressing the mbar/hPa key,
pull the air tube off the instrument!
The draft sensor is calibrated
(0.00 InW or hPa).

Carefully replace the air tube.
Wait until the measured value has
stabilized.

Record the measured draft value.

This value is stored with the current
measuring values.
The measuring mode is continued.

Attention:
No pressure values Exceeding

(

150 hPa mbar)!

Note:
For draft Measurement, connect air tube to the positive (+)
connector only.

±

60 inches of H O

2

±

Start draft measurement

16