Model ot-2 system maintenance 4 – Teledyne OT-2 - Oxygen transmitter system User Manual

Page 31

background image

4-3

Teledyne Analytical Instruments

Model OT-2 System

Maintenance 4

4. Place the new cell on the cap with the sensing surface

face down and the gold contact rings up.

5. Position the cell and cap so that the cell is inside the cell

block. Press up on the cap and rotate. Tighten the cap
screws.

4.2.1

Cell Warranty

The cell warranty is sensor specific. Please contact Teledyne

Electronic Technologies/ Analytical Instruments on explicit war-
ranty on sensors.

NOTE: Evidence of damage due to tampering or mishandling will

render the cell warranty null and void.

4.3

Scrubber Replacement

To ensure proper operation of the OT-2 System, it is recom-

mended to replace the scrubber every 6 months or more fre-
quently if the sensor is reaching end of life prematurely.

It is important that the replacement of the scrubber be com-

pleted within one hour of it's removal to prevent back diffusion of
oxygen from the atmosphere to the sensor.

During scrubber replacement, it is recommended that the 3-

way valve is switched to "Span Gas". This will minimize the system
recovery time.

NOTE: If the sensor is exposed to air through back diffusion, from

the atmosphere it may cause the sensor to re-stabilize
slowly.

4.4

Filter Servicing

To ensure the proper performance and sample flow, it is

recommended that the sample filter be inspected when the cell is
replaced. The filter requires both a 1" and a 1

1/16

" open end

wrench to remove the filter bowl. Turn off the sample flow valve
prior to servicing filter. The filter should be cleaned or replaced as
required.