3 Operation

Model OT-2 System

3-8

Teledyne Analytical Instruments

NOTE: This procedure is more critical on the 0-10 ppm range and

less critical on the 0-100 or 0-1000 ppm range.

3.3.2

Setting the Span Pot

1. Switch the system from zero gas to span gas and allow

time for stabilization.

Figure 3-6: Span Pot Location

NOTE: The span gas should have an O

2

concentration of 80% -

100% of full scale. For example, if the range is set at 0-100
ppm, the span gas should be a certified grade gas of
about 90 ppm O

2

in N

2

or Argon.

2. If 90 ppm gas is used, adjust the span pot (Figure 3-6)

to read 90.0 ppm on the LCD display. The output to
the Elsag Baily Total Flow

TM

computer will have a

signal of 4.6VDC (into a >10K ohm load).

The formula for this is:

Span Gas % F.S. x 4

Output in VDC

+ =

100

1

Display

Span

Span Pot