Teledyne LXT-280 User Manual
Page 43

Dissolved Oxygen Transmitter
Calibration Correction
Teledyne Analytical Instruments
33
Table 5-2: Relative Humidity Correction Factors
Temperature in °C
Relative
Humidity
0
5
10
15
20
25
30
35
0%
1.000 1.000 1.000 1.000 1.000 1.000 1.000 1.000
10%
.999
.999
.999
.998
.998
.997
.996
.994
20%
.999
.998
.998
.997
.995
.994
.992
.989
30%
.998
.997
.996
.995
.993
.991
.987
.983
40%
.998
.997
.995
.993
.991
.987
.983
.978
50%
.997
.996
.994
.992
.988
.984
.979
.972
60%
.996
.995
.993
.990
.986
.981
.975
.967
70%
.996
.994
.992
.988
.984
.978
.971
.961
80%
.995
.993
.990
.987
.982
.975
.967
.956
90%
.995
.992
.989
.985
.979
.972
.962
.950
100%
.994
.991
.988
.983
.977
.969
.958
.945
Table 5-3: Elevation Correction Factor
Elevation Correction Factor
Feet
0
1,000 2,000 3,000 4,000 5,000 6,000 7,000
0
1.000
.964
.930
.897
.865
.834
.804
.775
100
.996
.961
.926
.893
.861
.831
.801
.772
200
.993
.957
.923
.890
.858
.828
.798
.770
300
.989
.954
.920
.887
.855
.825
.795
.767
400
.986
.950
.916
.884
.852
.822
.792
.764
500
.982
.947
.913
. 880
.849
.819
.788
.761
600
.978
.943
.910
.877
.846
.816
.787
.758
700
.975
.940
.906
.874
.843
.813
.784
.756
800
.971
.937
.903
.871
.840
.810
.781
.753
900
.968
.933
.900
.868
.837
.807
.778 .750
- 1220 - Multipoint flammable gas and vapor detection system (50 pages)
- 212R - Thermal conductivity analyzer (28 pages)
- 235 - Thermal conductivity analyzer (38 pages)
- 275R - Portable turbine generator purge gas analyzer (21 pages)
- 2000A-EU - General purpose thermal conductivity analyzer (86 pages)
- 2000XTC - Thermal conductivity analyzer (40 pages)
- 2010A - Split architecture thermal conductivity analyzer (110 pages)
- 2010B - Split architecture thermal conductivity analyzer (98 pages)
- 2020 - Explosion proof thermal conductivity analyzer (80 pages)
- 2120 - Trace Nitrogen in Argon Analyzer (66 pages)
- 2120XL - Trace Nitrogen Analyzer (85 pages)
- 2230R - Process Hydrogen Analyzer (26 pages)
- 2240 – Portable Handheld Hydrogen Leak Detector, 3rd generation (updated 1/31/11) (30 pages)
- 2240 - Portable Handheld Hydrogen Leak Detector, 3rd generation (revision 2/29/08) (40 pages)
- 2240 – Portable Handheld Hydrogen Leak Detector, 2nd generation (13 pages)
- 2750 - Portable turbine generator gas analzyer (40 pages)
- 300P - Percent oxygen analyzer (24 pages)
- 306WA - Analog trace oxygen analyzer (46 pages)
- 311 - Portable trace oxygen analyzer (19 pages)
- 311D - Portable trace oxygen analyzer with digital meter (18 pages)
- 311XL - Portable trace oxygen analyzer (18 pages)
- 316RA / RB / RAD / RBD - Oxygen analyzers (24 pages)
- 319R - Oxygen analyzer (23 pages)
- 320 Series - Portable oxygen detectors (24 pages)
- 326, 327 and 328 - Oxygen analyzers (45 pages)
- 329R - Oxygen analyzer (22 pages)
- 335 - Analog control room monitor for personnel safety (24 pages)
- 356WA - Analog trace oxygen analyzer (42 pages)
- 3000MA - Paramagnetic oxygen analyzer (63 pages)
- 3000MA - Paramagnetic oxygen analyzer Addendum (2 pages)
- 3000MB - Paramagnetic oxygen analyzer (59 pages)
- 3000PA - General purpose percent oxygen analyzer (69 pages)
- 3000PAEU - General purpose percent oxygen analyzer (78 pages)
- 3000PB - Bulkhead mount percent oxygen analyzer (82 pages)
- 3000TA - General purpose trace oxygen analyzer (75 pages)
- 3000TA-EU - General purpose trace oxygen analyzer (89 pages)
- 3000TA-XLEU - Trace oxygen analyzer (108 pages)
- 3000TB - Bulkhead mount trace oxygen analyzer (78 pages)
- 3000TB-XL - Trace oxygen analyzer (78 pages)
- 3000ZA - Trace oxygen analyzer (81 pages)
- 3000ZA-3X - Trace oxygen analyzer (72 pages)
- 3000ZA2G - Zirconium oxide analyzer (72 pages)
- 3000 Ultra Trace - PPB oxygen analyzer (72 pages)
- 3010MA - Paramagnetic oxygen analyzer, includes 0-100% range (88 pages)
- 3010MA – Paramagnetic oxygen analyzer, no 0-100% range – (superceded) (88 pages)