Section 2: Sensor and Sampling Techniques
page 5
diameter of tubing, number of interconnecting fittings, dead 
space in tubing and manifolds; will influence the measure-
ment value and response time.
The high capacitance HTF™ sensors can be installed either 
directly in the line to be sampled (in-situ), or in a slip stream 
of a sample system (extractive).
To assure a long and accurate performance of the sensor, it 
should be protected from contaminants such as liquids 
(water, oil etc.), and particulates. The sintered stainless steel 
sensor encapsulation protects from particulates larger than 
100 microns, finer particulates (e.g. from degraded desic-
cant or rust) should be filtered with a particulate filter with 
suitable capability, do not use hygroscopic filter materials.
2.3.1 In-situ Installation
In-situ installation is recommended only for measurements 
where the gas pressure is expected to vary little, the gas is 
expected to be free of contaminants, the gas temperature is 
within the operating specifications of the sensor, and there is 
no chance of liquids coalescing. Examples of applications 
suited for in-situ installations are: pure gases, output of des-
iccant dryers (for instrument air), glove boxes, etc. For most 
other applications in-situ installation should be avoided for 
the following reasons:
•
Sample conditioning is almost always necessary to 
avoid exposure of the sensor to liquid water and other 
contaminants, such as hydrocarbons, which may dam-
age the sensor or affect accuracy over time.
•
Variations in line pressure affect the reading of the sen-
sor because dewpoint varies with pressure.
•
If the gas line is under pressure, it is more likely that 
water condensation occurs which may damage the sen-
sor.