Calibration adjustment: in this mode the user c – Teledyne 8800T - Trace moisture analyzer - transmitter User Manual
Page 30

8800T User’s Manual
page 26
tainer is labeled with the proper values.
5. Calibration Adjustment: In this mode the user can
enter a sensor specific Adjustment Value to improve the 
instruments’ accuracy in the range of -50°C to -10°C, 
this adjustment has very little or no effect outside this 
range. The Calibration Adjustment Value is printed on 
the shipping container supplied with the sensor, or may 
be derived if the sensor can be exposed to a known ref-
erence. 
① Deriving an Adjustment Value: If the adjustment 
value for your sensor is not known, contact your repre-
sentative and follow the instructions in paragraph B on 
the next page. If the value is not available, then it may 
be derived using a manual method, provided that a 
moisture calibration lab is available.
The calibration must be done in the range of -25°C to 
-20°C (attempting to calibrate outside this range may 
cause inaccuracies), the dewpoint must be kept stable 
during calibration and it must be measured accurately 
by a reference instrument such as a chilled mirror. One 
must also note the state of pressure compensation and 
turn it on or off as applicable to the reference.
The necessary steps are described below (also refer to 
the Set-Up State flow diagram in Appendix B):
a. Make sure that the low attenuation and low dewpoint 
are correctly entered for this sensor.
b. SpanCheck™ the sensor.
c. Dry-down the sensor for at least 12 hrs.
d. Expose the sensor to the known dewpoint (-25°C to 
-20°C), for a sufficient time (at least 30 min.)
e. Go to the 
&$/$'- mode and press the UP button, the
display will show the current value, in the range of -1.99 
to +1.99 followed by blinking horizontal lines.
