Teledyne Analytical Instruments

Total Hydrocarbon Analyzer

Appendix

IF USED OUTSIDE OF THESE SPECIFICATIONS
THE INSTRUMENT MAY BE DAMAGED AND
HAZARDOUS CONDITIONS MAY BE PRODUCED.

Appendix

Specifications

Ranges: 0-10, 0-100, 0-1000ppm hydrocarbon stan-

dard (methane equivalent)
(0-1ppm low range is optional.)

Sensitivity: 1% of full scale

Accuracy: ±2% of full scale at constant temperature

(Applies to standard ranges.)

Response Time: 90% in less than 15 second

Operating Temperature: 41°F to 110°F (5°C to 43.3°C)

Fuel: Mixture of 40% hydrogen, 60% nitrogen

recommended
(Other mixtures are possible. Contact TET/AI
for applications assistance.
)

Sensor Type: Flame Ionization Detector with Flame Guard

Signal Output: Internal - Digital Panel Meter

External - 0-1VDC, 4-20mA DC Isolated

Power Requirement: 115VAC ± 10% 50/60 Hz

230VAC ± 10% 50/60Hz

90VA nominal, 150VA maximum

Enclosure: 19" (48.3cm) Rack-Mounted Steel Enclosure

19" (48.3cm) W x 8.75" (22.2cm) H x 15.5"
(39.4cm) D