4.3

Preliminary Electronic Adjustment

After the foregoing preliminary checks and adjustments have been

made, turn the analyzer and recorder (if applicable) switches on (see Figure
4-1) and make the following checks and adjustments:

4.3.1

Meter-Recorder Alignment

If the application involves the use of an auxiliary recording and/or

indicating device, the following procedure should be used to check and
adjust the output circuit so that the meter and recorder are in close agree-
ment:

1. Set the analyzer span control so that the dial indicates the

number recorded in the Application Data sheet in the Appendix
under “Start-up Span Setting”.

2. Place the selector switch in the "LO" position.

3. Adjust the door-mounted zero control until the meter stabilizes

at a full scale indication.

4. Check the recorder reading. If the recorder’s input sensitivity

coincides with the analyzer’s output specification (see
Appendix: Specifications), the recorder should be at a balance
at its full scale marker.

Figure 4-1: Front Panel of Model 402R-EU Total Hydrocarbon Analyzer