fuse holder will slide out. There are two fuses in use and are visible in
the clip.

4. Remove the bad fuse and replace it with a 5x20mm 0.5 A,

250 VAC, IEC time lag (T) fuse (P/N F1128).

5. Replace the fuse holder into its receptacle, pushing in firmly until it

clicks.

5.2

Sensor Installation or Replacement

5.2.1 When to Replace a Sensor

On the trace analyzers, the Micro-fuel Cell will typically fail due to exces-

sive zero offset caused by loss of water. Large zero offsets will result in inaccu-
rate SPAN settings. To test for this condition, purge the sensor and sample
system with ZERO gas (typically nitrogen with less than 1 ppm O

2

). It may take

several hours for the sample system and cell to return to a stable low level of O

2

if it has been exposed to higher levels of O

2

. If the instrument will not come

down to an acceptably low reading on zero gas, then a depleted cell or a sample
system leak should be suspected. Typically offsets of 1.0 ppm or less are
acceptable. A Cell failure may also be indicated by an inability to SPAN, or
slow response to changes in O

2

concentration at levels below 100 ppm.

CAUTION: Read the section

Cell Warranty Conditions, below, before replac-

ing the cell.

CAUTION:

After replacing the Micro-Fuel Cell, the analyzer must be
recalibrated. See

Calibration in chapter 4.

5.2.2 Ordering and Handling of Spare Sensors

To have a replacement cell available when it is needed, TAI recommends

that one spare cell be purchased when the current cell 's warranty period is
approximately two thirds over.

CAUTION: Do not stockpile cells. The warranty period starts on the day of

shipment. For best results, do not order a new spare cell to soon.

The spare cell should be carefully stored in an area that is not subject to

large variations in ambient temperature (75 °F nominal), and in such a way as to
eliminate the possibility of incurring damage.

CAUTION: Do not disturb the integrity of the cell package until the cell is to

actually be used. If the cell package is punctured and air is permit-
ted to enter, cell-life will be compromised.