Effect of CO2 on B-2 Cell Life

1. This curve assumes continuous exposure.
2. Intermittent exposure will extend life. In

general, the CO

2

effect is cumulative and

the average CO

2

concentration should be

used to find the predicted cell life.

3. Cells can be used to make spot check

measurements of O

2

in the presence of

high concentrations of CO

2

(up to 50%

and more). Any sample containing CO

2

should be purged as soon as a constant
reading is obtained.

4. Abnormally slow response and recovery

is characteristic of cells used as indicated
in notes 1, 2, and 3 above.

5. The reduction in cell life is primarily due

to a drop in output. Cell life will reduce
to a point where the instrument can no
longer be spanned. Instruments using B-2
cells in CO

2

atmospheres should be

spanned at intervals of 10-20% of the
predicted cell life.

6. The CO

2

effect on B-2 cells is

independent of the O

2

level. Usage in

CO

2

contributes significantly to cell life

reduction and thus affects cell warranty.

Figure 4-2: Effects of CO2 on B-2 Cell Life