TELEDYNE ANALYTICAL INSTRUMENTS

Oxygen Alarm Monitor

Appendix

Appendix

Specifications

Range: 0-25% Oxygen

Sensitivity: 0.5% of full scale

Accuracy: ±2% of full scale (at constant temperature)

±5% of full scale (over operating temperature
range once the system has equilibrated)

Response Time: 90% in less than 15 seconds at 25°C (B-3 cell)

Operating Temperature: 15-45°C

Reproducibility: ± 1% of full scale

Sensor Type: B-3, B-5, C-3

Readout: High resolution 4

1

/

2

" analog meter

Battery Life: 48 hours (non-alarm condition)

Power Requirements: 115VAC @ 50/60 Hz charges and maintains on

trickle charge two NiCad battery packs.

Enclosure: Dust-tight steel enclosure (wall mountable, two

brackets top and bottom, each with two

5

/

16

"

holes, 8" center-to-center).
11.5"(H) x 9"(W) x 4.125"(D)
(22.9cm x 29.2cm x 10.4cm)

Weight: 11.7lbs (5.3Kg)

Alarms: Factory Set:

CAUTION = 20.0%
DANGER

= 19.5%

Audible Horn
Visual Red Indicator Lamps