Orion Instruments Jupiter 2xx User Manual
Page 29
46-648 Jupiter Magnetostrictive Transmitters
29
Unit with Display
Communicates
Error
Displayed?
No Level Signal
Error Codes
No
No
Yes
Reload all
Parameters
Yes
Yes
Yes
Corrupt Display
Text?
Replace Electronic
Module
No
Yes
Replace Electronic
Module
Same Error
Message?
Retest
No
Yes
Out of Calib
Analog Loop
Correct?
Trim 4-20 mA
No
Yes
No
Call Factory
Display Reading
Correctly?
4
Yes
No
Unit with Display
Communicates
3
Check probe
length
Float is not
detected
Consult Factory
Check Offset
and
4-20 mA
Set Points