2 status messages – Orion Instruments Jupiter 2xx User Manual
Page 26
26
46-648 Jupiter Magnetostrictive Transmitters
2.3.2
Status Messages
Display Message
Action
Comment
OK
None
Normal operating mode
Initial
None
Shown at power-up during
self check
TrimReqd
Factory set Loop values are defaults,
Consult Factory
loop output may be in accurate
Cal Reqd
Factory set default calibration parameters
Consult Factory
are in use, level reading may be inaccurate
Lo Temp
Present temperature in electronics
Transmitter may need to be moved
compartment is below -40° F (-40° C)
to ensure temperature is within
specification
Hi Temp
Present temperature in electronics
Transmitter may need to be moved
compartment is above +176° F (+80° C)
to ensure temperature is within
specification
Float 2
No level signal detected from float 2
Make sure 2 floats are being used,
Fail
are not damaged, and within
measuring range
Float 1
No level signal detected from float 1
Make sure float is not damaged
Fail
and within measurement range
No Signal
No signal detected from any float
Make sure float is not damaged
and within measurement range
LoopFail
Loop current differs from expected value
Consult Factory
Snsr Brd
Measurement board not responding
Consult Factory
DfltParm
Internal non-volatile parameters have been
Consult Factory
reset to default values