beautypg.com

Appendix 3: calculating temperature coefficients, Appendix 4: meter factory default settings – Nova-Tech pH-CON 300 User Manual

Page 26

background image

51

50

14. Appendix 3: Calculating Temperature Coefficients

To determine the temperature coefficient of your sample solution use this formula:

C

T2

- C

T1

TC = 100 x _________________________

C

T1

(T

2

- 25) - C

T2

(T

1

- 25)

TC = Temperature coefficient

C

T1

=Conductivity at Temp. 1

C

T2

= Conductivity at Temp. 2

T

1

= Temp. 1

T

2

= Temp. 2

25 = 25°C

NOTE: A controlled temperature water bath is ideal for this procedure.

1.

Immerse the probe into a sample of your solution and adjust the temperature
coefficient to 0% (that is, no compensation) by performing the following:
A. From conductivity measurement mode, press the SETUP key.
B. Press the

▲ key until the lower display reads P8.0.

C. Press the ENTER key twice. The lower display reads tCO and the upper

display shows the temperature coefficient value.

D. Press the

▼ key until the upper display shows 0.0.

E. Press ENTER key to confirm the value.
F. Press CAL/MEAS key twice to return to measurement mode.

2.

Wait for 5 minutes. Note T

1

and C

T1

(conductivity at T

1

).

3.

Condition the sample solution and probe to a temperature (T

2

) that is about 5°C

to 10°C different from T

1

, and note the conductivity reading C

T2

.

NOTE: Record your results for future reference. Ideally T

1

and T

2

should bracket

your measurement temperature, and should not differ by more than 5°C.

4.

Calculate the temperature coefficient of your solution according to the formula
shown above.

5.

Enter the temperature coefficient you calculated into the meter.
A. From conductivity measurement mode, press the SETUP key.
B. Press the

▲ key until the lower display reads P8.0.

C. Press the ENTER key twice. The lower display reads tCO and the upper

display shows the temperature coefficient value (should be 0, per step 1-D

above).

D. Press the

▲ key until the upper display shows your calculated temperature

coefficient.

E. Press ENTER key to confirm the value.
F. Press CAL/MEAS key twice to return to measurement mode.

The calculated temperature coefficient will now be applied to all the meter readings.

15. Appendix 4: Meter factory default settings

Resetting the meter to factory default settings clears all calibration data and most
other setup functions you might have changed. The following settings will remain as
you have set them:

• temperature unit of measure (°C or °F)
• The temperature offset calibration value

NOTE: Conductivity/TDS and pH data are cleared separately from each other.
• To clear pH data, see Sub group P4.0 on page 31
• To clear conductivity/TDS data, see Sub group P9.0 on page 40

Type

Parameter

Default

Remarks

pH parameters

P1.1

Viewing pH

– – –

No calibration data for pH 1.68

P1.2

calibration data

– – –

No calibration data for pH 4.01

P1.3

– – –

No calibration data for pH 7.00

P1.4

– – –

No calibration data for pH 10.00

P1.5

– – –

No calibration data for pH 12.45

P2.1

View electrode offset

0.00 mV

No offset adjustment

P2.2

View electrode slope

100.0%

No slope adjustment

P3.1

Ready indicator

Ready on

Ready indicator on; auto endpoint off

P3.2

# pH cal. points

3

3 pH calibration points available (1-5 point range)

P3.3

°F or °C

no default

°F or °C remains as selected

P4.0

pH factory default

no

Retains your current settings

Conductivity/TDS parameters

P5.1

Viewing con./TDS

– – –

No cal. data for R1 (0.00-19.99 µS/0.00-9.99 ppm)

P5.2

calibration data

– – –

No cal. data for R2 (0.0-199.9 µS/10.0-99.9 ppm)

P5.3

– – –

No cal. data for R3 ( 0-1999 µS/100-999 ppm)

P5.4

– – –

No cal. data for R4 (0.00-19.99 mS/1.00-9.99 ppt)

P5.5

– – –

No cal. data for R5 (0.0-199.9 mS/10.0-199.9 ppt)

P6.1

Viewing con./TDS

1.000

No offset for R1 (0.00-19.99 µS/0.00-9.99 ppm)

P6.2

calibration data

1.000

No offset for R2 (0.0-199.9 µS/10.0-99.9 ppm)

P6.3

1.000

No offset for R3 ( 0-1999 µS/100-999 ppm)

P6.4

1.000

No offset for R4 (0.00-19.99 mS/1.00-9.99 ppt)

P6.5

1.000

No offset for R5 (0.0-199.9 mS/10.0-199.9 ppt)

P7.1

Ready indicator

Ready on

Ready indicator on; auto endpoint off

P7.2

°F or °C

no default

°F or °C remains as selected

P7.3

ATC on or off

ATC on

P7.4

Set TDS factor

0.5

Adjustable from 0.4 to 1.0

P8.1

Temp. coefficient

2.1% per °C

Adjustable from 0 to 10% per *C

P8.2

Normalization temp.

25°C

Adjustable from 15 to 30°C / 59 to 86°F

P9.0

factory default

no

Retains your current settings

http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682