P3.0: ph measurement configuration – Nova-Tech pH-CON 300 User Manual

Page 15

background image

29

28

From measurement mode

1.

Press the Mode key to select pH
measurement mode.

2.

Press Setup key to enter Set Up mode.

3.

Press the

and keys to scroll through

subgroups until you view parameter P3.0.

See figure

4.

Press the ENTER key twice to select
parameter 3.2.

See figure

5.

Press the

and keys to select 2, 3, 4,

or 5 point pH calibration.

6.

Press the ENTER key to confirm selection
and to proceed to step 5 of P.3.3. Press the
CAL/MEAS key twice to return to meas-
urement mode.

SETUP

pH

READY

SETUP

ON

C

C

D

D

P3.2: Selecting number of pH calibration points

Program P3.2 lets you select the number of pH calibration points the meter will use
in calibration mode: 2, 3, 4, or 5. The meter will automatically exit calibration mode
after you have calibrated to your selected number of points.

From measurement mode

1.

Press the Mode key to select pH
measurement mode.

2.

Press Setup key to enter Set Up mode.

3.

Press the

and keys to scroll

through subgroups until you view
parameter P3.0.

See figure

4.

Press the ENTER key to select
parameter 3.1.

See figure

5.

Press the

and keys to select the

configuration you require.
•OFF switches the READY indicator off.
•ON switches the READY indicator on.
•ON and HOLD together switches the

auto endpoint feature on.

6.

Press the ENTER key to confirm selec-
tion and to proceed to step 4 of P.3.2.
Press the CAL/MEAS key twice to return
to measurement mode.

P3.0: pH measurement configuration

P3.1: READY indicator and auto endpoint function

This program lets you select:

“READY indicator on” to indicate when the reading is stable.

“READY indicator off” for faster meter response.

Auto endpoint function on. Select auto endpoint on to “hold” the reading when

it is stable for more than 5 seconds. The display automatically freezes, and the
HOLD indicator appears on the left side of the display. Press the HOLD key to
release the display and access other functions.

SETUP

pH

READY

SETUP

ON

READY

SETUP

OFF

READY

HOLD

SETUP

ON

6.4

A

A

B

B

Notes

Meter default is set for Ready indicator on,
and auto endpoint function off.

You can also change the Ready indicator and
auto endpoint function in Program P7.1
(available from conductivity/TDS mode).
Any changes you make to the Ready indica-
tor/ auto endpoint function in pH mode will
also change in conductivity mode.