Specifications, Error messages – Nova-Tech pH-CON 300 User Manual
Page 23

10. Specifications
Mode
pH
Temperature
Conductivity
TDS
0 to 19.99 µS
0.00 to 9.99 ppm
–2.00 to
0 to 100°C /
0 to 199.9 µS
10.0 to 99.9 ppm
Range
16.00 pH
32.0 to 212.0°F
0 to 1999 µS
100 to 999 ppm
0 to 19.99 mS
1.00 to 9.99 ppt
0 to 199.9 mS
10.0 to 199.9 ppt
0.01 µS, 0.1 µS, 0.01 ppm, 0.1 ppm,
Resolution
0.01 pH
0.1°C or °F
1 µS, 0.01 mS,
1ppm, 0.01 ppt,
0.1 mS
0.1 ppt
Accuracy
±0.01 pH
±0.5°C / ±0.9°F
±1% full scale
±1% full scale
up to 5 points
offset in 0.1°
up to 5 points
up to 5 points
Calibration
(pH 1.68, 4.01,
increments
(one point
(one point
7.00, 10.01, 12.45)
up to 5°
per range)
per range)
pH slope: 80 to 110%
Conductivity cell constant: K = 1.0
Conductivity/TDS temperature coefficient: adjustable from 0.0 to 10.0% per °C
Conductivity/TDS normalization temperature:
adjustable from 15 to 30°C / 59 to 86°F
Conductivity-to-TDS conversion factor: adjustable from 0.4 to 1.0
Temperature compensation: automatic or manual from 0 to 100°C / 32 to 212°F
Display: dual LCD
Operating temperature: 0 to 50°C / 32 to 122°F
Power: four 1.5 V AAA batteries (included)
Battery life: > 200 hours continuous use
Probe connectors:
pH/mV: BNC connector
Conductivity/TDS/Temperature: notched six-pin connector
Dimensions:
Meter: 7.5"L x 3.5"W x 1.75"H
(19.1 cm x 8.9 cm x 4.5 cm)
Boxed: 9.2"L x 9.2"W x 2.75"H
(23 cm x 23 cm x 7 cm)
Probe: 6.8"L x 1.3" dia (17.3 cm L x 3.2 cm dia)
Weight:
Meter: 1.0 lb (0.5 kg)
Boxed: 2.0 lbs (0.9 kg)
45
44
9. Error Messages
LCD
Indicates
Cause
Solution
Display
Err
Unrecognized input
Wrong input in
Release key. Select
annunciator
from keypad
selected mode.
valid operations
depending on mode.
CAL & Err
Calibration error
Wrong value input
Check your input value,
annunciators
at calibration.
clean probe.
blink
Dirty probe.
See Calibration sections or
Probe Maintenance section.
Battery
Low battery level
Need new batteries
Clean battery contacts.
indicator blinks
or battery
Replace batteries with fresh
connection is bad.
ones, noting polarity.
If error persists, or the meter shows incorrect values, return the meter.
See “Warranty” and “Return of Items” on page 52.
For a complete diagram of the display, see page 4.