Troubles

Battery symbol

on LCD

Check lists

Countermeasures

“Lo” on screen

(result below

32.0

℃ or 89.6℉)

“Hi” on screen

(result below

42.9

℃ or 109.3℉)

No response/

Automatic

reset when pulling

out insulator

Battery drained
Battery in wrong polarity?

Poor battery contact

Change batteries

Take out battery, reinsert

battery correctly

Low battery

Change new battery

Thermometer correctly angled

for the eardrum / forehead?

Follow user’s manual to

correctly angle for the

eardrum / forehead.

Please check the operation

method

Follow user’s manual for

proper measurement.

“Errp” on screen

“Errp” on screen

“Err5” on screen

Hardware problem
Environment temperature

beyond the

scope of

10℃~40℃(50.0℉~104.0℉)

for Ear

15℃~40℃(59.0℉~104.0℉)

for Forehead

Place the thermometer

into environment

temperature 10℃~40℃

(50.0℉~104.0℉) for Ear

15℃~40℃ (59.0℉~

104.0℉) for Forehead

Operation error, More than 15

seconds without test result in

measure heart rate mode

please follow the

instruction in user manual

“Er25” on screen

Thermometer temperature

measurement circuit failure

Consult manufacturer

Consult manufacturer

TROUBLESHOOTING GUIDE

21

NOTES:
1. Under an environment with electrostatic discharge, the unit may

malfunction and may require user to reset the unit.
2. Dispose with care; please consult your retailer or manufacturer

for instructions, if needed.