Mobi Technologies 70120 DualScan ULTRA User Manual

Page 17

background image

Immunity test

IEC 60601

test level

Compliance level

Electrostatic

discharge (ESD)

lEC 61000-4-2

±

6 kV contact

±

8 kV air

Electrical fast

transient/burst

IEC 61000-4-4

±

2kV for power

supply lines

±

1kV for input/

output lines

±

1kV

differential

mode

±

2kV

common mode

Voltage dips,

short

interruptions

and voltage

variations on

power supply

input lines

IEC 61000-4-11

The device is intended for use in the electromagnetic environment specified

below. The customer or the user should assure that it is used in such an

environment.

Surge

IEC 61000-4-5

Electromagnetic

environment - guidance

±

6 kV contact

±

8 kV air

Floors should be wood,

concrete or ceramic tile.

If floors are covered with

synthetic material, the

relative humidity should

be at least 30 %.

Not applicable

Mains power quality

should be that of a typical

commercial or hospital

environment.

Not applicable

Mains power quality

should be that of a typical

commercial or hospital

environment.

<5% UT

(>95% dip in UT)

for 0.5 cycle

40% UT

(60% dip in UT)

for 5 Cycles

70% UT

(30% dip in UT)

for 25 Cycles

<5% UT

(>95% dip in UT)

for 5 sec

Mains power quality

should be that of a typical

commercial or hospital

environment. If the user

of the device requires

continued operation

during power mains

interruptions, it is

recommended that the

device is powered from

an uninterruptible power

supply or a battery.

Not applicable

Power frequency

(50/60Hz)

magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic

fields should be at levels

characteristic of a typical

location in a typical

commercial or hospital

environment.

NOTE: UT is the a.c. mains voltage prior to application of the test level.

Guidance and manufacturer's declaration - electromagnetic immunity