Record Mode 
 
The M1 can store up to 240 pressure measurements (also called 
values or points) in a single record (REC) session. For maximum 
flexibility, a REC session can be: 
• Automatic - the current value is automatically stored every 5 
seconds, for up to 20 minutes
• Manual - the current value is stored every time the U key is
pressed, up to 240 times
 
Both types of REC sessions can store between 1 and 240 
measurements. The measurement data is preserved in non-volatile 
memory until another REC session is started. 
 
 
REC Data: 
 
In NORMAL mode enter REC Data mode by pressing both the B 
and F keys until the REC icon begins flashing. 
 
Second, choose the type of REC session. 
 
To start an automatic REC session, press the F key. This will 
delete the previously stored pressure measurement and store the 
current pressure measurement. Then, every 5 seconds, the current 
pressure value will be stored. 
 
To start a manual REC session, press the U key. This will delete 
the previously stored session and store the current pressure 
measurement. Then, every time the U key is pressed, the current 
pressure value will be stored. 
 
For both types of REC session, the pressure value on the LCD will 
blank briefly every time a sample is stored, and after 240 
measurements have been recorded, the following step (stop and 
exit REC Data mode) will happen automatically. 
 
Third, to stop and exit REC Data mode, press both the B and F 
keys. All previously stored REC session data will be saved and the 
M1 will return to NORMAL mode (the REC icon will go off). This 
step can be done anytime the REC icon is flashing. 