Appendix – Meriam 4150X ATEX I.S. Certified HART Communicator User Manual
Page 33
29
APPENDIX
PRODUCT SPECIFICATIONS
BASE UNIT: 10.375" L, 4.25” W – top, 4.375" W – middle, 1.875” D. Handle: 2.50” W, 1.25"D
ABS plastic case with shock absorbing boot, 1.7 lbs including batteries and boot
DISPLAY: 128 x 128 pixel graphic display, 2.6” x 2.6” viewable area, backlight
POWER: six (6) AA alkaline batteries, six (6) AA NiMH batteries or 100 – 240 Vac universal power adapter (P/N
Z9B000007) connected to the 4150 / 4150X via the Communication / AC Adapter Dongle.
Note: Use only approved AA alkaline batteries in Intrinsically Safe units. DO NOT USE unapproved
AA alkaline batteries, AA NiMH batteries or other types of AA batteries in Intrinsically Safe units. Do
not use the Dongle or AC Adapter with Intrinsically Safe units in a potentially hazardous environment.
See the “Hazardous Area Use” section of this manual, the specifications below and the Intrinsic Safety
Control Drawing in the Appendix for more information.
BATTERY LIFE (Approximate): 60 hours using AA alkaline batteries without backlight
27 hours using AA alkaline batteries with backlight
65 hours using NiMH alkaline batteries without backlight
28 hours using NiMH alkaline batteries with backlight
TEMPERATURE SPECIFICATIONS:
Storage: -40° F to 140° F (-40° C to 60° C)
Operating: 23° F to 122° F (-5° C to 50° C)
CONNECTIONS: Standard banana jacks (¾” center) for HART communications
Communication / AC adapter connector for Dongle connection: AC power adapter and serial cable
(to PC port) connects to Dongle; Dongle connects to 4150 / 4150X.
APPROVALS: 4150X (ATEX/I.S.)
ATEX Intrinsically Safe II 1 G; EEx ia IIC T4
c
UL
us
Listed
Intrinsically Safe, Exia
Class I, Div. 1 Groups A, B, C, D: T4
-5º C < Ta < +50º C
Entity parameters:
Li = 0, Ci = 0
Approved Batteries for ATEX/I.S.: (6) AA alkaline Duracell PC1500
(6)
AA
alkaline
Duracell
MN1500
(6) AA alkaline Varta 4906
APPROVALS: 4150 (general purpose)
MEMORY – separate, dedicated, non-volatile flash memory for 4150 / 4150X firmware, DOFs, and Configurations
DOF Memory: 560 HART devices (based on average DOF size)
Configuration Memory:
200 HART devices (based on average configuration size)
AUTO SHUT-OFF - occurs after user-selected period of keypad inactivity or can be disabled (both 4150 / 4150X auto
shut off and backlight auto shut off are user selectable).
DIAGNOSTICS – 4150 / 4150X self-tests upon power up. Advisories are posted on the display.