Dsca, High performance signal conditioners, Dsca specifications – Impulse DSCA34 User Manual
Page 2: Ordering information

High Performance Signal Conditioners
DSCA
Visit our website www.dataforth.com
137
DSCA
Specifications
Input Range Limits
-200°C to +850°C (100
Ω
Pt)
-80°C to +320°C (120
Ω
Ni)
Input Protection
Continuous
240Vrms max
Transient
ANSI/IEEE C37.90.1
Sensor Excitation Current
≈
250µA
Lead Resistance Effect
+0.02°C/
Ω
Output Range
See Ordering Information
Load Resistance (I
OUT
)
600
Ω
max
Current Limit
8mA (V
OUT
), 30mA (I
OUT
)
Output Protection
Short to Ground
Continuous
Transient
ANSI/IEEE C37.90.1
CMV, Input to Output, Input to Power
Continuous
1500Vrms max
Transient
ANSI/IEEE C37.90.1
CMV, Output to Power
Continuous
50VDC max
CMR (50Hz or 60Hz)
160dB
Accuracy
See Ordering Information
Conformity
±0.025% (100
Ω
Pt)
±0.07% (120
Ω
Ni)
Adjustability
±3% zero and span
Stability
Input Offset
±1µV/°C
Output Offset
±6ppm/°C (V
OUT
), ±20ppm/°C (I
OUT
)
Gain
±60ppm/°C
Output Noise, 100kHz Bandwidth
250µVrms (V
OUT
), 1µArms (I
OUT
)
Bandwidth, 3dB
3Hz
NMR
85dB at 60Hz, 80dB at 50Hz
Response Time, 90% Span
250ms
Open Input Response
+ Lead
Upscale
Lead
Non-deterministic
x Lead
Downscale
Power Supply
Voltage
15 to 30VDC
Current
25mA (V
OUT
), 55mA (I
OUT
)
Sensitivity
±0.0001%/%
Protection
Reverse Polarity
Continuous
Transient
ANSI/IEEE C37.90.1
Environmental
Operating Temp. Range
40°C to +80°C
ATEX Group II, Category 3
20°C to +40°C
Storage Temp. Range
40°C to +80°C
Relative Humidity
0 to 95% Noncondensing
Emissions EN61000-6-4
ISM, Group 1
Radiated, Conducted
Class A
Immunity EN61000-6-2
ISM, Group 1
RF
Performance A ±0.05% Span Error
ESD, EFT, Surge, Voltage Dips
Performance B
Mechanical Dimensions
2.95" x 0.89" x 4.13"
(h)(w)(d)
(75mm x 22.5mm x 105mm)
Mounting
DIN EN 50022 -35x7.5 or -35x15 rail
Module
DSCA34
Typical at T
A
=+25°C and +24V supply voltage
Ordering Information
Model
100
Ω
Ω
Ω
Ω
Ω
Pt **
DSCA34-01
-100°C to +100°C
2, 3, 4
±0.08%
±0.16°C
(-148°F to +212°F)
DSCA34-02
0°C to +100°C
2, 3, 4
±0.10%
±0.10°C
(+32°F to +212°F)
DSCA34-03
0°C to +200°C
2, 3, 4
±0.08%
±0.16°C
(+32°F to +392°F)
DSCA34-04
0°C to +600°C
2, 3, 4
±0.05%
±0.30°C
(+32°F to +1112°F)
DSCA34-05
-50°C to +350°C
2, 3, 4
±0.05%
±0.20°C
(-58°F to +662°F)
120
Ω
Ω
Ω
Ω
Ω
Ni **
DSCA34N-01
0°C to +300°C
2, 3, 4
±0.15%
±0.45°C
(+32°F to +572°F)
Input Range
Output
Accuracy
1
Range
**
RTD Standards
Type
Alpha Coefficient
DIN
JIS
100
Ω
Pt
0.00385
DIN 43760
JIS C 1604-1989
120
Ω
Ni
0.00672
Output Range
Part No. Suffix
Example
1. -10V to +10V
NONE
NA
2. 0V to +10V
NONE
DSCA34-01
3. 4 to 20mA
C
DSCA34-01C
4. 0 to 20mA
E
DSCA34-01E
Output Ranges Available
NOTES:
(1) Includes conformity, hysteresis and repeatability.
Installation Notes:
1.) This Equipment is Suitable for Use in Class I, Division 2, Groups A, B, C, D, or Non-Hazardous Locations Only.
2.) Warning - Explosion Hazard - Substitution of Components May Impair Suitability for Class I, Division 2.
3.) Warning - Explosion Hazard - Do Not Disconnect Equipment Unless Power Has Been Switched Off or The
Area is Known to be Non-Hazardous.