Installation, Installation guidelines, Mechanical installation – Hach-Lange FILTERTRAK FT 660 sc User Manual User Manual
Page 9: Instrument mounting, Installation guidelines mechanical installation
Installation
D A N G E R
Multiple hazards. Only qualified personnel must conduct the tasks
described in this section of the document.
Installation guidelines
N O T I C E
High internal temperatures will damage the instrument components.
• Do not operate the instrument in direct sunlight or next to a heat
source.
• Install the instrument indoors in a clean environment with minimum
vibration.
Note: Particulate contamination from the surrounding area is a major cause of
error.
• Install the instrument level.
• Keep the flow rate and operating temperature as constant as possible
for best performance.
Mechanical installation
Instrument mounting
Install this instrument on a flat, vertical surface. Refer to the illustrated
steps in
. Mounting hardware (two ¼-20 bolts) is supplied by the
customer.
Install the instrument as near to the sampling point as possible to
minimize lag time of the sample flow.
Refer to
for instrument dimensions and clearance
specifications.
Note: As an alternative, this instrument can be installed on the optional floor stand.
Figure 3 Instrument mounting
English 7