beautypg.com

How to clean spills, Troubleshooting, Errors – Hach-Lange FILTERTRAK FT 660 sc User Manual User Manual

Page 22: Warnings, Errors warnings

background image

How to clean spills

C A U T I O N

Chemical exposure hazard. Dispose of chemicals and wastes in
accordance with local, regional and national regulations.

1. Obey all facility safety protocols for spill control.
2. Discard the waste according to applicable regulations.

Troubleshooting

Problem

Possible cause

Resolution

Low (< 1) or negative
readings

Head assembly is not
correctly installed on the
body of the instrument

Correctly install the head
assembly on the body of the
instrument. If not, the red
status light will flash.

Laser failure

Contact Technical Support.

PMT/Electronics failure

Errors

When an error occurs:

• The reading on the measurement screen flashes.
• All outputs are held.
• The red status light on the instrument turns on.

To access the current instrument errors, select Diagnostics, [Select
Sensor], Error List.

Select an error and confirm to identify the cause of the error. Refer to

Table 2

.

Table 2 Error list

Error

Possible cause

Resolution

A/D FAILURE

A/D conversion failure

Contact Technical Support.

LASER FAILURE

Laser did not work correctly
and is turned off

FLASH FAILURE

Data log and event log not
working

LOW SIGNAL
ERROR

Signal is too low (less than
3.0 mNTU)

Make sure that the head
assembly is installed
correctly. If not, the red
status light will flash.
Contact Technical Support.

Warnings

When a warning occurs:

• A warning icon flashes and a message is shown on the bottom of the

measurement screen.

• The red status light on the instrument starts flashing.

Warnings do not affect the operation of menus, relays and outputs.
To access the current instrument warnings, select Diagnostics, [Select
Sensor], Warning List.
Select a warning and confirm to identify the cause of the warning. Refer
to

Table 3

.

20 English